WHO malaria nucleic acid amplification test external quality assessment scheme: results of eleven distributions over 6 years
Malaria Journal volume 24, Article number: 94 (2025)
Abstract
Background
The World Health Organization (WHO) recommends parasite-based diagnosis of malaria before treatment. The use of nucleic-acid amplification (NAAT) for detection of Plasmodium spp. has expanded rapidly in recent years, for epidemiological research globally and clinical care in high-resource settings. Data from NAATs are frequently used to inform policy decisions, so quality control is essential to ensure results are reliable and comparable. Therefore, robust quality control, including an external quality assessment (EQA) scheme targeting malaria NAATs, is essential. The WHO Global Malaria Programme and the UK National External Quality Assessment Service (UK NEQAS) have collaborated since 2017 to implement a global malaria NAAT EQA scheme.
Methods
Panels of specimens containing five major species of human-infecting Plasmodium at various parasite concentrations and negative samples were created in lyophilized blood (LB) and dried blood spot (DBS) formats. Two distributions per year were sent, containing five LB and five DBS specimens. Samples were validated by expert referee laboratories prior to distribution. Between 37 and 51 laboratories participated in each distribution and submitted results online. Participants were scored based on their laboratory's stated capacity to identify Plasmodium species, and individual laboratory reports were sent which included performance comparison with anonymized peers. Change in performance over time was calculated using a generalized mixed model with a logit link function.
Results
Participating laboratories were located in 42 countries. Sample format (DBS or LB) and parasite density were found to significantly affect performance, while referee labs performed better at identifying P. falciparum samples than non-referee labs. Performance of laboratories improved significantly over time, especially for lower density and P. falciparum samples.
Conclusions
Results from the first eleven distributions indicate that the EQA scheme has facilitated improved performance of laboratories over time, highlighting the value of implementing such programmes. EQA schemes are critical to safeguarding the reliability of data and diagnoses, especially in situations where NAAT methodologies and protocols are used. In future, funders should make participation in an EQA scheme a requirement for laboratories, and countries can take initiatives to embed such schemes into their own national assessment programmes.
Background
Malaria remains a significant challenge to global health, with 608,000 deaths from the disease in 2022, the majority occurring in sub-Saharan Africa [1]. Accurate and timely diagnosis is crucial for effective case-management of Plasmodium infection and elimination efforts in malaria-endemic countries, and since 2010 the World Health Organization (WHO) has recommended confirmation with parasite-based diagnosis of malaria before treatment is given [2]. The main tools for diagnosing malaria in endemic countries are microscopy and antigen-detecting malaria rapid diagnostic tests (RDTs).
The use of nucleic acid amplification-based tests (NAATs) for detection of Plasmodium infection has expanded significantly in recent decades. Since the identification of Plasmodium 18S ribosomal RNA [3] and the development of the first polymerase chain reaction (PCR) for Plasmodium by Snounou and colleagues in 1993 [4], numerous more advanced and/or modified NAAT methods have been developed, such as nested, multiplex and real-time PCR, reverse-transcriptase PCR, loop-mediated isothermal amplification, nucleic acid sequence-based amplification assays and cluster regularly interspaced short palindromic repeats (CRISPR)-based detection [5,6,7,8,9,10,11]. NAATs are commonly used in malaria research studies such as clinical trials for new anti-malarial medicines or vaccines, drug efficacy studies, as a reference standard for evaluations of new diagnostics, or for epidemiological studies [12, 13]. Their use for diagnosis in high-income countries is also increasing. NAATs have a lower limit of detection than other assays (down to as few as 1–100 parasites/mL of whole blood in some cases) [8, 14,15,16].
Due to their very high sensitivity, NAATs are attractive tools to consider for use in low transmission countries moving towards malaria elimination, where the majority of infections are of very low parasite density and often below the limit of detection of RDTs and microscopy [17]. If Plasmodium falciparum and Plasmodium vivax are both endemic, but largely undetectable by non-NAAT methods, NAATs could be used to identify and enable appropriate treatment in elimination efforts such as mass screen and treat programmes [18].
As well as detecting the presence of Plasmodium and species identification, NAATs have several other uses such as to study presence of molecular markers including those conferring resistance to anti-malarials, or to assess whether P. falciparum parasites have deletions in their pfhrp2/3 genes, which can lead to erroneous results on the most commonly-used RDTs [19]. NAATs allow for detection of mixed or multiclonal infections, whereas RDTs often cannot detect mixed infections and both they and microscopy are not capable of detecting multiclonal infections. Whole genome sequencing can be used to conduct diversity studies, for example of surface proteins, and can be used to discover medically important variants such as single nucleotide polymorphisms [20], or to distinguish between recrudescence and reinfection [21, 22]. Reliable and accurate NAAT assays are needed to support assessments of efficacy of new malaria vaccines in development and in the early stages of use; of new therapeutics and to assess accuracy of new diagnostics; along with supporting epidemiological understanding of malaria and informing how and where to strengthen surveillance efforts. Thus, data from NAATs are frequently used to inform policy decisions for vaccines, medicines and diagnostics, so quality control is essential to ensure the results generated by NAATs are reliable and also comparable between the laboratories conducting them. Quality assessment schemes can facilitate improvement in laboratory performance and in harmonizing methodologies [23]. While other diagnostic techniques, such as microscopy and RDTs, have their own versions of quality assurance schemes [24, 25], EQA schemes for malaria NAATs are rare.
The WHO Global Malaria Programme and the UK National External Quality Assessment Service (UK NEQAS) Parasitology collaborated to launch the WHO Malaria NAAT EQA scheme in January 2017, targeting laboratories in low and high malaria transmission and resource settings to promote reliability and comparability of molecular data. Panels of Plasmodium samples are shipped to participating laboratories twice a year, and after analysis and submission of results the laboratories can assess their performance over time and in relation to other participating laboratories, although all results by laboratory remain confidential.
Results of the first three distributions of this scheme were published by the same authors in 2020 [26]. Those results showed that the type of sample [dried blood spots (DBS) or lyophilized blood (LB)], Plasmodium species, and being a referee laboratory were significant factors in performance. As the scheme has progressed, the additional data obtained have allowed for more in-depth analysis of changes in performance of laboratories over time and other factors affecting performance. This paper presents data from the first eleven distributions of the scheme along with challenges faced by laboratories and scheme coordinators.
Methods
Enrolment of laboratories
In late 2014, reference and research laboratories in all WHO regions were approached and asked to complete a survey regarding malaria NAAT activities and invited to join the scheme. In order to identify as many eligible laboratories as possible, laboratory networks supporting clinical trials and malaria activities were approached. Awareness of the scheme is continually being increased through a range of mechanisms such as presentations at international conferences and meetings, and creation and maintenance of a WHO webpage featuring relevant documentation and contact information. Donors funding research that is reliant on molecular methods are encouraged to provide the budget and require participation in the scheme. As such, laboratories have continued to join the scheme at every distribution since the beginning.
For the first six distributions, participation was free. In 2020, a tiered fee system was introduced, based on the location (high/ low-income country) and type of laboratory (private, reference, or research laboratory).
EQA source material
Details of the source materials and preparation of the EQA samples have been published previously [26]. Briefly, leftover clinical samples consisting of EDTA-anti-coagulated peripheral blood were diluted in whole blood regarded as Plasmodium negative by UK NHS Blood and Transplant (UKNHSBT), within 48 h of receipt from the diagnostic laboratory. Pre-dilution parasite densities were determined by expert microscopists from UK NEQAS Parasitology, counting the number of parasitized cells in a sample size of 10,000 red blood cells on thin blood films to obtain a percentage parasitaemia and converted to parasite density (number of parasites per µL) using the red cell count. For samples included in the first distribution, red cell counts of 5 × 1012 per litre was assumed, while red cell counts in subsequent distributions were determined in the initial, pre-dilution samples using a C-Chip DHC-N01 Disposable haemocytometer (NanoEnTek Inc. via MT Promedt Consulting GmbH, Germany). For cultured parasites, a thin blood film was made from the undiluted culture and a haemocytometer was used from the outset to count erythrocytes in all cases, in order to obtain the pre-dilution parasite density in parasites per µL of synchronized ring-stage parasites. For all samples, clinical and cultured, dilution to the desired parasite density was performed using parasite-negative whole blood supplied by UK NHSBT. All positive and negative samples were confirmed by PCR at multiple referee laboratories after sample production.
Panel composition
Each panel consisted of Plasmodium positive and negative samples and included five DBS samples, containing 50 µL of blood per spot, and five LB samples, containing 500 µL of blood per vial. Concentration of samples ranged from 0.018 to 1,100,000 parasites/µL (Table 1). Not every species was included in each distribution.
Paired samples were included in six distributions. These samples had identical species and concentrations, in DBS and lyophilized blood forms, to enable direct comparison of performance between these two formats. Eleven pairs were included, totalling 22 samples.
Panel distribution
The first panel of samples was shipped in January 2017. Panels were shipped roughly every six months, with the 11th distribution being sent in September 2022. Most laboratories received samples sent at ambient temperature via courier. In some cases, samples were sent via air freight to the nearest airport for collection.
Referee laboratories
Seven laboratories were selected to be referee laboratories, based on their experience and publication record of using a range of NAATs and location of their laboratory [27]. These labs conducted molecular analysis on samples prior to distribution to the wider group of participating labs, for confirmation of parasite content and concentration. Referee labs were blinded to the intended results during their referee role as well as during the later panel testing phase. Referee laboratories were unaware in which future distributions the samples would be included, and different sample IDs were used for the EQA scheme distributions. Six of these laboratories also participate in the EQA scheme.
Data reporting and analysis
After each shipment of samples, a web portal hosted by UK NEQAS was open for six to eight weeks for participants to submit their results. Within 24 h of closing of each distribution, intended results were uploaded on the participant’s portal stating the actual contents of each sample. A detailed report stating the performance of the individual laboratories in that distribution and their scores was uploaded within four to six weeks of closing of each distribution.
Before enrolling into the scheme, laboratories were asked to provide information on what plasmodia they were able to detect to species and genus level, which were then compiled to create an individual laboratory profile. Apart from the initial results, results presented in this paper are based on results adjusted according to the laboratory profiles.
Data analysis apart from that based on submission number was performed using Stata15 (Stata Corp, College Station, USA). Differences in performance were assessed using design-based F-test.
Analysis of change of performance over time based on submission number
Since laboratories could join the scheme at any time, a new variable was created which indicated the submission number for each laboratory/submission combination. Laboratories which joined the scheme at the eleventh distribution will only have data for one submission (on distribution eleven), while laboratories that joined at the start of the scheme and participated in every distribution will have data for submission numbers 1 to 11. As an example, if a laboratory joined the scheme at the beginning but missed three distributions and had submitted results in eight distributions by distribution eleven, their highest submission number would be eight.
Fewer Plasmodium knowlesi, Plasmodium malariae and Plasmodium ovale samples had been included in the scheme by distribution eleven than the number of P. falciparum or P. vivax samples. Therefore, these three species have been combined into one group and analysed together to ensure sufficient data for modelling. Additionally, not every species, sample type and density combination were included in each distribution. Laboratory submission numbers cannot exceed the last distribution at which a particular species/sample format/density combination was used. For instance, P. falciparum DBS samples with < 100 parasites/µL were only included in distributions 1 and 4, meaning laboratory performance data can only have submission numbers 1 to 4 for this combination. The range of submission numbers varied from 3 to 11, dependent on the species/sample format/density combination. A density threshold of 100 parasites/µL was used for this analysis, as opposed to 2 parasites/µL used for the previous analysis, as there were too few samples of below 2 parasites/µL to enable robust modelling.
The change in performance in terms of correctly identifying a sample within a laboratory was assessed using a generalized mixed model with a logit link function, laboratory as the cluster variable and submission number as the independent predictor. To utilize all the data, but not extrapolate beyond the available data, separate models were created for each species/sample format/density combination. Only combinations used at least once after the third distribution are modelled (i.e. the range of submission numbers must exceed three). Analysis was conducted using the GAMLj module in Jamovi (Gallucci, M. (2019). GAMLj: General analyses for linear models. [jamovi module]. Retrieved from https://gamlj.github.io/; The jamovi project (2021). jamovi. (Version 1.8) [Computer Software]. Retrieved from https://www.jamovi.org.).
Results
Number of laboratories participating
Overall, 75 laboratories had enrolled in the scheme and submitted results by the eleventh distribution. While panels were sent to between 45 and 66 laboratories in each distribution, the number submitting results ranged from 37 to 51 over the eleven distributions (Table 2). This number is continually changing—new laboratories enrol after each distribution, laboratories cannot always participate in every distribution for various reasons, and occasionally laboratories have to drop out. The number of samples with results submitted ranged from 324 to 479. The number of laboratories submitting results for lyophilized blood was higher than those submitting DBS results in every distribution.
Participating laboratories were located in 42 countries. Twenty-one laboratories (28%) were located in 13 African countries, 17 (23%) were in 11 countries in Asia, and eight laboratories (11%) were in six countries in Europe. Fourteen laboratories (19%) were located in nine countries in South and Central America, 12 (16%) in the two North American countries, and three laboratories (4%) were located in Australia.
All seventy-five laboratories submitted results in at least one distribution. Five laboratories submitted in only one distribution, while eleven submitted in all eleven distributions. However, as the scheme is constantly expanding and new laboratories join at each distribution, some of the laboratories which did not submit results in all distributions did not necessarily miss a distribution, but joined the scheme after it started. For example, of the twelve laboratories which submitted results three or fewer times, seven did not join the scheme until distribution nine or later and so all of these except one had submitted in every distribution since enrolling. Twenty-two laboratories submitted results in all the distributions since they joined the scheme, while by the eleventh distribution, nine laboratories had only missed one distribution since joining the scheme.
Laboratory profile
All laboratories report being able to detect P. falciparum, while eight laboratories detect only P. falciparum and no other species. The majority of laboratories (67, 89%) report being able to detect P. falciparum and P. vivax, with or without other species. Plasmodium knowlesi is the least common species detected, with 29 laboratories (39%) able to detect this species, while 57 (76%) and 55 (73%) laboratories report being able to detect P. ovale and P. malariae, respectively.
Performance
Across all distributions, laboratories performed best on negative samples, with 90.6% of negative samples being identified correctly (Fig. 1). Among positive samples, P. falciparum was the most likely to be identified correctly, at 89.1%. As all laboratories are able to identify P. falciparum samples, there was no difference between the raw and adjusted results of these samples. When looking at raw results, without taking into consideration the laboratory profile, 80.2% of P. vivax samples were correctly identified, 60.6% of P. ovale samples, 48.7% of P. malariae samples, and P. knowlesi was the most difficult to identify, with only 36.7% identified correctly. However, when results are adjusted according to the laboratory profiles, the percentage of correctly identified P. knowlesi samples increased to 75.4%, 67.1% of P. malariae samples were correctly identified, as were 86.9% of P. vivax samples.
Referee laboratories
Overall, there was no significant difference in performance between the six referee laboratories that participate in the scheme and non-referee laboratories (86.9% and 84.0%, p = 0.65) (Fig. 2). However, when broken down by species, referee laboratories showed significantly better performance at identifying P. falciparum samples (99.0% and 88.3% correctly identified in referee and non-referee laboratories, respectively, p < 0.001). There was no significant difference among any other species, negative samples, or format of sample overall. When broken down by parasite density, referee labs showed markedly better performance than non-referee labs against samples ≤ 2 parasites/µL (94.1% and 70.5%, p = 0.002).
Performance by sample format and parasite density
Performance on analysis of lyophilized blood was more accurate than on DBSs, with 88.2% of lyophilized samples and 79.9% of DBSs correctly identified (p < 0.01) (Fig. 3). Among the paired samples (meaning the samples had identical species and concentrations) included in six of the distributions, accuracy was 91.6% for lyophilized samples and 77.7% for DBSs (p < 0.01).
When samples were broken down into parasite density, higher density samples were better detected overall and on each sample format, where data was available. There was strong evidence to show better performance at densities above 2 parasites/µL, by DBS and lyophilized blood, and among P. falciparum, P. vivax and P. knowlesi samples (Fig. 4). This biggest difference in performance was seen among P. falciparum samples, with 68.9% and 94.5% being correctly identified for samples ≤ 2 parasites/µL and > 2 parasites/µL respectively. No comparison was possible for P. malariae and P. ovale. Difference in performance was especially marked among DBSs below and above 2 parasites/µL, with only 50.0% of DBS samples below 2 parasites/µL being correctly identified compared with 79.7% of those above this threshold (p < 0.01).
Adjusted accuracy of external quality assessment results by parasite concentrationa. *Denotes difference with a p-value below 0.05. DBS: Dried Blood Spots. LB: lyophilized blood. aResults adjusted for laboratory capacity. bAll samples of P. malariae and P. ovale were > 2 parasites/µL. Three P. ovale sample was pooled and therefore concentration was unknown and not included in the analysis
Analysis method used
The most commonly used nucleic acid extraction method, using silica columns, accounted for 60.8% of the samples analysed, and the remaining samples were analysed using seven other methods (Additional file 1). As the majority of labs were using a similar method, it is difficult to draw conclusions about which methods showed the best or worst performance (Additional file 2). The most commonly used amplification method was real-time single target PCR (34.8%), followed closely by nested PCR (33.0%) (Additional file 1). There was no significant difference in performance by method of nucleic acid amplification (Additional file 2).
Performance of laboratories over time
Negative samples
The odds of correctly identifying a parasite negative sample were significantly higher for lyophilized samples compared to DBS (Odds Ratio (OR) 1.618; 95% CI 1.029–2.545; p = 0.037). There was no significant relationship with submission number (p = 0.596), with consistently high odds of correctly detecting negative samples.
Plasmodium falciparum samples
The odds of correctly identifying a sample containing P. falciparum parasites were influenced by submission number for samples below 100 parasites/µL, for both DBS and lyophilized blood samples (Fig. 5, Additional file 3). Data were available for four submissions for lower density DBS samples and 11 submissions for higher density samples. For each increase in submission number, the odds of correctly identifying one of these samples below 100 parasites/µL increased by 2.70 for DBS samples (95% CI 1.57–4.65, p < 0.001) and 1.34 for lyophilized blood samples (95% CI 1.15–1.56, p < 0.001). The change in performance was not significant (p > 0.355) for higher density samples, as performance of these samples was consistently high from the beginning of the scheme.
Plasmodium vivax samples
For each additional submission made by a laboratory, the odds of correctly identifying a P. vivax sample increased by 1.28 for lower density DBS samples (95% CI 1.06–1.54, p = 0.010), over 11 submission distributions (Fig. 5, Additional file 4). The change in performance was not significant for lyophilized blood samples of either higher (p = 0.072) or lower density (p = 0.739), and there were insufficient data to model DBS samples above 100 parasites/µL.
Plasmodium knowlesi , P. malariae and P. ovale samples
For samples containing P. knowlesi, P. malariae or P. ovale the odds of a correct result increased by 1.37 (95% CI 1.12–1.66, p = 0.001) by submission number for DBS samples over 100 parasites/µL, while the change in performance of lower density DBSs was not found to be affected by submission number (p = 0.235) (Fig. 5, Additional file 5). Performance was shown to improve for lower density lyophilized samples, with odds of correctly identifying a sample increasing by 1.24 (95% CI 1.24–1.39, p < 0.001), however while the trend for the higher density lyophilized samples was the same, these results were not statistically significant (p = 0.093).
Challenges and reasons for non-participation
Despite being enrolled in the scheme for several distributions, some laboratories were unable to participate in every distribution, or had to stop participation. Discussions with laboratory personnel indicated that this inconsistency or discontinuation of the scheme was due to a number of reasons, some of which were outside the control of the laboratory personnel. For example, obtaining import permits in time for importing samples into the country has been problematic for some laboratories since the beginning of the scheme, inability to pay customs fees, availability of reagents, and lack of laboratory staff for the testing can be challenging. The COVID-19 pandemic exacerbated many of these challenges and caused additional struggles for laboratories due to the shutdown of transport routes, laboratory staff time being diverted to other projects and inability of staff to access the laboratories. Departure of the UK from the European Union has led to disruptions for some laboratories located in Europe, as import permits are now required for European laboratories, where they were not required previously.
The majority of laboratories have been able to pay or have had their fees paid by third-parties since the introduction of fees in 2020. On several occasions where a laboratory has been unable to continue due to funding constraints, the WHO has covered the cost of participation. These fees do not cover WHO costs of running the scheme and, therefore, external sources of funds have been mobilized and will continue to be required in the future.
Discussion
High quality NAAT for malaria diagnosis is essential to support research that informs policy and the development of new medicines, diagnostics and vaccines, as well as strategies to move towards malaria elimination in some settings. Having an EQA scheme like the one presented here can support laboratories to improve their performance and maintain it over time.
Prior to enrolment, laboratories submitted a profile outlining which Plasmodium species they could detect to species and genus level. While all laboratories could detect P. falciparum and the majority detected P. vivax, detection of the other species was much less common, and the performance at detecting these samples was poorer than for P. falciparum or P. vivax. While P. falciparum and P. vivax are the two main species infecting people in malaria-endemic countries, it may become increasingly important to detect the less common species using NAATs, as they are less easily detected by RDTs, and high-quality microscopy, which can differentiate between species, is difficult to maintain.
Plasmodium knowlesi is now endemic in all Southeast Asian countries except Timor-Leste, and is the most common cause of malaria in some regions [28]. There is no good alternative to NAATs to detect this species as the ability of RDTs to detect P. knowlesi is poor [29] and there is no RDT that specifically targets this species or discriminates it from others. Misidentification of P. knowlesi as another species by microscopy is due to its similarity to other species (e.g. to P. malariae, and to P. falciparum at the ring stage) [30]. Therefore, good quality NAATs to detect this species are essential and the lack of a well-performing assay could be a serious hindrance to malaria case management and control in endemic countries.
False-negatives were more common than false-positive results. Some results showed misidentification of species, so while the presence of Plasmodium nucleic acids was detected, the correct species was not reported. The consequences of either type of error, or of misidentifying a species, can be critical, not only in clinical practice, but also in certain types of research such as drug efficacy studies.
Performance on samples of lyophilized blood was markedly higher than DBS samples, likely due to the larger volume of blood (500 µL) which can be extracted from lyophilized samples compared to DBS (50 µL), thus increasing the chance of capturing parasite-derived nucleic acids. Comparisons of identical paired samples showed 14 percentage points better accuracy on lyophilized blood than DBSs. The accuracy for DBS samples with parasite densities < 2 parasites/µL was only 50.0% in the scheme. Two parasites/ µL is considered the threshold that NAAT assays should be consistently reaching to offer significant performance advantage over microscopy and RDTs. DBSs are very useful tools for analysing archived samples and are much easier to collect and transport than lyophilized samples, but care must be taken when analysing these types of samples as the limit of detection is higher than for lyophilized samples because the tested sample volume for DBS is usually lower than for lyophilized samples. A study has shown that the concentration of Plasmodium DNA extracted from DBS is much lower than from whole blood, resulting in lower detection of malarial DNA after PCR [31], while another study has shown that when comparable volumes of blood are used in DBS and whole blood format, there was only a moderate reduction in DNA amplification from DBS samples compared to whole blood [32]. Therefore, it must be understood that negative results on DBS samples may not necessarily mean absence of Plasmodium parasites, but too few to detect by NAATs on that specimen type.
Performance over time
The performance of laboratories improved significantly over time, even after adjusting for concentration and sample format, especially for lower density and P. falciparum samples. Performance of these samples was generally poorer in the first submissions, and improved rapidly over time participating in the scheme, to become more similar in performance to that on higher density samples. Performance of negative samples and higher density P. falciparum samples was very high to begin and has been maintained. As the overall performance has improved, it has started to plateau, so it is now important to help poorer performing laboratories to improve their accuracy, as well as supporting the higher performing laboratories to maintain their performance over time. Establishing a network of laboratories to facilitate higher performing laboratories to support those with lower scores can help to raise the performance of these laboratories. If a laboratory is shown to have poor performance, they can request a repeat panel to be shipped for re-testing, and can be put in contact with an expert who can provide remote technical support, or with other laboratories from the network that might have faced similar challenges and/or are using similar methodologies.
Performance was not shown to be affected by methodology or type of nucleic acid detected. Rather than recommending that laboratories change their methods to another method of extraction, improving performance of the method they are using already is more efficient at improving performance.
While the overall number of laboratories participating in the scheme has increased, the proportion participating in each distribution has remained about the same, with several laboratories in each distribution not submitting results. While some of the factors leading to missed participation in a distribution are outside the control of the lab, it is important to discuss with lab personnel why this is happening, what the main barriers are, and ways in which the coordinators of the scheme can facilitate more consistent participation in the scheme. Factors such as issues with funding or availability of reagents can be problematic. Where laboratories are not able to submit results in time, in exceptional circumstances, the window to submit has been extended. While the scheme was originally free for participants, fees were introduced in distribution seven. Many laboratories are still unable to pay these fees, and need to be encouraged to build the budget for these fees into funding proposals. Several labs did withdrawal participation after introduction of fees.
Adapting the scheme to track biological threats
The development of such a robust EQA scheme allows for expansion into assessing performance on other aspects of P. falciparum infection. Since distribution five, random P. falciparum samples have been included which have pfhrp2 and/or pfhrp3 gene deletions. These gene deletions have been detected in all malaria endemic regions at varying prevalence and are able to evade detection by the most commonly used (i.e., HRP2-detecting) RDTs [33]. Surveillance efforts to monitor these gene deletions have expanded over the last decade, and their confirmation relies entirely on NAATs. As confirmation of these deletions relies on the absence of a DNA product, it is an especially challenging technique. Including these samples in the panels will allow those laboratories which test for the presence or absence of these genes to assess their ability to identify correctly pfhrp2 and/or pfhrp3 gene deletions among P. falciparum samples.
An addition to the EQA scheme will be included in future distributions, with P. falciparum samples with/without anti-malarial (AM) drug resistance markers, to assess laboratories’ ability to detect a range of currently known drug resistance markers. A pilot phase including PfK13 single nucleotide polymorphisms, which confer partial resistance to artemisinin drugs, and other partner drug resistance markers is planned for 2025. Laboratories who wish to participate in the AM resistance marker testing scheme will be sent their usual panel of samples for EQA testing, along with an additional panel with P. falciparum samples with/ without AM resistance markers. Adding this new scheme onto the existing EQA scheme saves on logistic and shipping costs and benefits both the participating laboratories and the coordinators of the scheme.
Inclusion of paired samples will also allow laboratories to assess their performance of molecular correction, an important tool used to differentiate recrudescence with new infections for interpreting therapeutic efficacy study outcomes.
Future role of NAAT
The COVID-19 pandemic highlighted the role that NAAT, backed up by EQA schemes, can play in infectious diseases diagnostics [34]. The pandemic led to a great expansion in NAAT capacity in many countries [35, 36] and countries want to maintain and leverage this enhanced capacity for expansion of molecular diagnostics to other diseases. It is hoped that this will then lead to an increase in demand for the malaria molecular EQA scheme. The results from the first eleven distributions of this EQA scheme indicate that performance of laboratories has increased over time, showing the positive impact that EQA schemes can have on NAAT performance.
The scheme coordinators are aware that there are a large number of laboratories that perform malaria NAATs that are not yet participating in the scheme, so an ongoing activity is to determine how best to encourage them to join the scheme. Another malaria molecular scheme is run by UK NEQAS, using LB samples only and laboratories can decide which is more suitable for them, depending on their needs.
These results show that the goal of this EQA scheme has been achieved in that it appears to have facilitated improved NAAT performance of laboratories over time. It has also revealed areas where improvement is still needed, including detection of very low-density infections (< 2 parasites/µL) and minor species. It is also proving itself to be a strong backbone to efficiently expand to meet needs for monitoring proficiency at detecting emerging biological threats such as pfhrp2/3 deletions and anti-malarial drug resistance. Donors should make participation in an EQA scheme a requirement for laboratories they are financing to apply NAAT for surveillance or research purposes, and countries can take the initiative to embed such schemes into their own national assessment programmes, rather than the current system of vertical schemes. This will lead to improved performance and accountability of those laboratories that conduct NAATs to inform surveillance, disease management and research.
Availability of data and materials
No datasets were generated or analysed during the current study.
References
WHO. World malaria report 2023. Geneva: World Health Organization; 2023.
WHO. Guidelines for the treatment of malaria. 2nd ed. Geneva: World Health Organization; 2010.
McCutchan TF, de la Cruz VF, Lal AA, Gunderson JH, Elwood HJ, Sogin ML. Primary sequences of two small subunit ribosomal RNA genes from Plasmodium falciparum. Mol Biochem Parasitol. 1988;28:63–8.
Snounou G, Viriyakosol S, Jarra W, Thaithong S, Brown KN. Identification of the four human malaria parasite species in field samples by the polymerase chain reaction and detection of a high prevalence of mixed infections. Mol Biochem Parasitol. 1993;58:283–92.
Bell D, Fleurent AE, Hegg MC, Boomgard JD, McConnico CC. Development of new malaria diagnostics: matching performance and need. Malar J. 2016;15:406.
Murphy SC, Prentice JL, Williamson K, Wallis CK, Fang FC, Fried M, et al. Real-time quantitative reverse transcription PCR for monitoring of blood-stage Plasmodium falciparum infections in malaria human challenge trials. Am J Trop Med Hyg. 2012;86:383–94.
Han ET, Watanabe R, Sattabongkot J, Khuntirat B, Sirichaisinthop J, Iriko H, et al. Detection of four Plasmodium species by genus- and species-specific loop-mediated isothermal amplification for clinical diagnosis. J Clin Microbiol. 2007;45:2521–8.
Imwong M, Hanchana S, Malleret B, Renia L, Day NP, Dondorp A, et al. High-throughput ultrasensitive molecular techniques for quantifying low-density malaria parasitemias. J Clin Microbiol. 2014;52:3303–9.
Snounou G, Viriyakosol S, Zhu XP, Jarra W, Pinheiro L, do Rosario VE, et al. High sensitivity of detection of human malaria parasites by the use of nested polymerase chain reaction. Mol Biochem Parasitol. 1993;61:315–20.
Schneider P, Schoone G, Schallig H, Verhage D, Telgt D, Eling W, et al. Quantification of Plasmodium falciparum gametocytes in differential stages of development by quantitative nucleic acid sequence-based amplification. Mol Biochem Parasitol. 2004;137:35–41.
Xiu L, Li H, Hu Q, Zhang Y, Chen SB, Wang C, et al. A versatile microfluidic platform for malaria infection screening and Plasmodium species genotyping. EBioMedicine. 2023;98:104898.
Murphy SC, Shott JP, Parikh S, Etter P, Prescott WR, Stewart VA. Malaria diagnostics in clinical trials. Am J Trop Med Hyg. 2013;89:824–39.
Hanscheid T, Grobusch MP. How useful is PCR in the diagnosis of malaria? Trends Parasitol. 2002;18:395–8.
Britton S, Cheng Q, McCarthy JS. Novel molecular diagnostic tools for malaria elimination: a review of options from the point of view of high-throughput and applicability in resource limited settings. Malar J. 2016;15:88.
Chang M, Johnston S, Seilie AM, Hergott D, Murphy SC. Application of dried blood spot sample pooling strategies for Plasmodium 18S rRNA biomarker testing to facilitate identification of infected persons in large-scale epidemiological studies. Malar J. 2021;20:391.
Hodgson SH, Douglas AD, Edwards NJ, Kimani D, Elias SC, Chang M, et al. Increased sample volume and use of quantitative reverse-transcription PCR can improve prediction of liver-to-blood inoculum size in controlled human malaria infection studies. Malar J. 2015;14:33.
Okell LC, Ghani AC, Lyons E, Drakeley CJ. Submicroscopic infection in Plasmodium falciparum-endemic populations: a systematic review and meta-analysis. J Infect Dis. 2009;200:1509–17.
WHO. Guidelines for malaria. Geneva: World Health Organization; 2022.
Gamboa D, Ho MF, Bendezu J, Torres K, Chiodini PL, Barnwell JW, et al. A large proportion of P. falciparum isolates in the Amazon region of Peru lack pfhrp2 and pfhrp3: implications for malaria rapid diagnostic tests. PLoS ONE. 2010;5: e8091.
Akoniyon OP, Adewumi TS, Maharaj L, Oyegoke OO, Roux A, Adeleke MA, et al. Whole genome sequencing contributions and challenges in disease reduction focused on malaria. Biology. 2022;11:587.
Beshir KB, Diallo N, Sutherland CJ. Identifying recrudescent Plasmodium falciparum in treated malaria patients by real-time PCR and high resolution melt analysis of genetic diversity. Sci Rep. 2018;8:10097.
SnounouGaB HP. The use of PCR genotyping in the assessment of recrudescence or reinfection after antimalarial drug treatment. Parasitol Today. 1998;14:462–7.
Shrivastava J. Assessors assemble: the need for harmonised external quality assessment schemes for emerging diagnostic methodologies in the field of parasitology. Trans R Soc Trop Med Hyg. 2019;113:820–2.
WHO. Malaria rapid diagnostic test performance: results of WHO product testing of malaria RDTs: round 8 (2016–2018). Geneva: World Health Organization; 2018.
Kettelhut MM, Chiodini PL, Edwards H, Moody A. External quality assessment schemes raise standards: evidence from the UKNEQAS parasitology subschemes. J Clin Pathol. 2003;56:927–32.
Cunningham JA, Thomson RM, Murphy SC, de la Paz AM, Ding XC, Incardona S, et al. WHO malaria nucleic acid amplification test external quality assessment scheme: results of distribution programmes one to three. Malar J. 2020;19:129.
WHO. External quality assurance scheme for malaria nucleic acid amplification testing: operational Manua. Geneva: World Health Organization; 2018.
Lee WC, Cheong FW, Amir A, Lai MY, Tan JH, Phang WK, et al. Plasmodium knowlesi: the game changer for malaria eradication. Malar J. 2022;21:140.
Yerlikaya S, Campillo A, Gonzalez IJ. A systematic review: performance of rapid diagnostic tests for the detection of Plasmodium knowlesi, Plasmodium malariae, and Plasmodium ovale monoinfections in human blood. J Infect Dis. 2018;218:265–76.
Huber JH, Elliott M, Koepfli C, Perkins TA. The impact of emerging Plasmodium knowlesi on accurate diagnosis by light microscopy: a systematic review and modeling analysis. Am J Trop Med Hyg. 2023;108:61–8.
Strom GE, Moyo S, Fataki M, Langeland N, Blomberg B. PCR targeting Plasmodium mitochondrial genome of DNA extracted from dried blood on filter paper compared to whole blood. Malar J. 2014;13:137.
Murphy SC, Daza G, Chang M, Coombs R. Laser cutting eliminates nucleic acid cross-contamination in dried-blood-spot processing. J Clin Microbiol. 2012;50:4128–30.
Thomson R, Parr JB, Cheng Q, Chenet S, Perkins M, Cunningham J. Prevalence of Plasmodium falciparum lacking histidine-rich proteins 2 and 3: a systematic review. Bull World Health Organ. 2020;98:558–68.
WHO. Quality Assessment (EQA) Programme for National and Subnational laboratories. Geneva: World Health Organization; 2021.
Bashir IM, Al-Waleedi AA, Al-Shaibani SM, Rajamanar M, Al-Akbari S, Al-Harazi A, et al. Strengthening laboratories in response to outbreaks in humanitarian emergencies and conflict settings: results, challenges and lessons from expanding PCR diagnostic capacities for COVID-19 testing in Yemen. PLoS ONE. 2024;19: e0298603.
Khan MM, Tahoun MM, Meredith LW, Barakat A, Safi H, Hanifi AN, et al. Implementation and expansion of laboratory capacity for molecular diagnostics in response to COVID-19 and preparedness for other emerging infectious diseases in the Islamic Emirate of Afghanistan. Influenza Other Respir Viruses. 2023;17: e13210.
Acknowledgements
The authors would like the thank the participants of the EQA scheme for the use of the results from their submissions in this manuscript. Thank you to the UK NEQAS Microbiology team for assistance with the EQA provision; Vaibhavi Mistry and Dipali Pindoria for coordinating the logistics of distributions of the panels; NEQAS (Birmingham Quality) for IT support; the participating laboratories and those that served as referees for pre-distribution quality checks of all specimens; and Dr Don van Schalkwyk from the LSHTM for malaria parasite cultures. Thank you to Colin Sutherland from LSHTM and Nick Edwards from the University of Oxford for their input.
Funding
This work was supported using funds from the World Health Organization.
Author information
Authors and Affiliations
Contributions
JC, PC and JS designed and coordinated the study. RT, MG, PC and JS analysed the data. RT, JC, PC and JS drafted the original manuscript. JS and AS conducted laboratory work including EQA sample preparation. SM, MA, XD, SI, EL, NL, DM and SN provided expert advice on developing the EQA scheme. All authors provided comments on and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
Samples for EQA material were prepared from leftover clinical specimens from patients attending the Hospital for Tropical Diseases in London and from residual blood samples referred from other hospitals across the UK cultured parasites were used for P. knowlesi and for some P. falciparum specimens. Under the terms of the UK Human Tissue Act, 2004, Schedule 4, section 45, supplementary part 2, ethics approval to use such samples for EQA purposes is not required.
Consent for publication
Not applicable.
Competing interests
The authors declare no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Thomson, R.M., Cunningham, J.A., Gatton, M.M. et al. WHO malaria nucleic acid amplification test external quality assessment scheme: results of eleven distributions over 6 years. Malar J 24, 94 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12936-025-05282-0
Received:
Accepted:
Published:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12936-025-05282-0